The Beran 455 Integrated
TransCal™ provides, in one box, all of the measuring devices required to calibrate the majority of vibration transducers. The unit includes a power amplifier to provide drive to the user's shaker and two channels of matched signal conditioning amplifiers, one for the reference transducer and the other for the test transducer.

The sine correlation measurement technique is used in the Beran 455 to ensure that accurate, repeatable measurements are provided over a wide frequency range and with changing measurement environments.


 

455

 

Internal signal conditioning amplifiers provide interfacing to charge, voltage and ICP type devices. Additionally, an internal power amplifier provides drive to an external shaker or alternatively, connection to an external power amplifier-shaker system may be made.

Calibration procedures include frequency step at traceable frequencies, frequency sweep or amplitude step - or all three if required. All user configuration and result display is via an in-built graphical interface.



 
MEASUREMENT UNCERTAINTIES
(complete system)
0.7% Reference Transducer
5Hz to 5kHz
±1.05%
2% Reference Transducer
5Hz to 5kHz
±2.15%
at an ambient temperature of 20° ± 10°C
 

The Beran 455 Integrated TransCal™ is fully compatible with the Beran 485 Application Software. This provides compatibility with Microsoft Windows packages and other databases. The 485 Application Software also includes a design module to allow users to customise calibration certificates.